Z-1899-2020: Varian Multileaf Collimator [MLC]; Millenium Multileaf Collimator [52, 80, 120 leaf]; High Definition 120

Recall record Z-1899-2020 for Varian Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1899-2020
Company in FDA record
Varian Medical Systems, Inc.
FDA record creation date
May 5, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

After a recent upgrade to the collimator software version 8.5, the firm became aware that the multi-leaf collimator leaves did not move during an arc treatment. The issue is related to a transient carriage primary- secondary interlock due to carriage fault on the MLC that immediately preceded the initiation of the arc treatment. The issue will occur only with MLC software version 8.5 and only affects conformal arc treatments, including VMAT and RapidArc. There have been no reports of adverse health consequences due to this issue.

Source facts

Action
Urgent Medical Device Correction Field Safety Notice letters were sent to customer beginning August 12, 2019. Letters were sent to U.S. and Canadian customers via Fed-Ex, and via local third-party courier or email for the rest of the world. Users were instructed to: 1. monitor MLC leaf movement on the 4DITC console screen, 2. observe the start of every arc treatment and verify that the leaves are moving as intended, and 3. stop treatment immediately if MLC leaves are not moving and contact Varian to assess the MLC status. Users may contact Varian Support for help to verify whether carriage faults have been recorded for the MLC. If carriage fault is confirmed in the TerminalLog file, users were instructed to contact Varian Service Representative to assess the MLC status. Customers were requested to complete and return the response form provided to returnresponse@varian.com, and to contact their local Varian Medical Systems service manager with any questions. The notification letter includes Varian Support Information with the following contact information: phone number for North America Customers: 1 888.827.4265; Global email address support@varian.com; and website contact address for customers outside of North America: https://www.myvarian.com/s/contactus?lang=en.
Initiated
2019-08-12
Root cause
Software Design Change
Terminated
2021-01-21
Distribution
Worldwide distribution - US Nationwide distribution including in the states of TN, FL, IN, PA, NV, WA, DC, CA, AL, IL, MD, VA, TX, MN, AR, GA, OR, MS, MO, NY, SC, DE, AZ, NH, MI, NC, OH, PA and OK. The country of Canada.
Recall number
Z-1899-2020
Recall status
Terminated
CDRH record ID
180396
FDA record created
2020-05-05
Recall event number
85269

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗