Skip to content

Public FDA record

Z-1899-2024: Brand Name: EsophaCap Product Name: EsophaCap (20mm diameter, 20 pores/inch) Model/Catalog Number: EC20T2

Recall record Z-1899-2024 for Lucid Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1899-2024
Company in FDA record
Lucid Diagnostics, Inc.
FDA classification date
May 24, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The EsophaCap sponge may be at increased risk of detaching from the string during removal of the device from the patient, particularly at the lower esophageal sphincter at the gastroesophageal junction or the upper esophageal sphincter between the pharynx and the esophagus, which is the narrowest part of the alimentary canal.

Source facts

Action
On April 24, 2024 direct customers were sent notification letters via email with recall instructions. Effectiveness Checks: The signed Acknowledgment and Receipt forms must be returned along with any product that is still in-hand at the clinical sites. Customer Follow-up: Follow-ups will be conducted via phone and email for customers who do not respond. Returned Product: Returned product will be put into quarantine until all recalled product is under our control. Access is limited to authorized personnel only. It will then be destroyed. Destruction: Devices will be cut and disposed of. Once the string is cut, the device will no longer be functional to use. PD1 Quarantined: 115 PD2 Quarantined: 99 Preventive Action Taken by Firm: All EsophaCap product in the field will be returned to Lucid Diagnostics. The EsophaCap product is no longer manufactured for the clinical studies. The last manufactured product was February 2023.
Initiated
2024-04-24
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of MN, MD.
Recall number
Z-1899-2024
Recall status
Open, Classified
CDRH record ID
207346
FDA classified
2024-05-24
Recall event number
94510

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗