The product has a software problem in which previous patient measurement data gets associated with another patient's image.
Source facts
Action
The firm, SIEMENS, issued a "CUSTOMER SAFETY ADVISORY NOTIFICATION" letter to all customers. The letter described the product, problem and actions to be taken. The customers were instructed to ensure that the first image in Exam Review is from the current patient; ensure that the Heart Cycle measurement tool is not selected on the Cale menu when entering Exam Review; be aware of this issue as described in the letter and perform four (4) or less measurements of the same measurement label.
Should you have any questions, please call (650) 694-5398.
Initiated
2009-01-14
Root cause
Software change control
Terminated
2011-04-06
Distribution
Worldwide distribution.
Recall number
Z-1900-2011
Recall status
Terminated
CDRH record ID
80551
FDA classified
2011-04-06
Recall event number
51497