Open original FDA record ↗Recall record
- FDA identifier
- Z-1901-2018
- Company in FDA record
- LivaNova USA
- FDA classification date
- May 18, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Firm is notifying customers that the Deep-Cleaning Service is now available in the United States. Stockert Heater-Cooler System 3T devices are suspected of microbial contamination and may contain visible biofilm.
Source facts
- Action
- On 04/20/2018, a Medical Device Correction notice was sent to customers informing them to return devices less than ten years old. Returned devices will receive a deep cleaning service, which will reduce microbial contamination to HPC < 100 CFU/ml of water and NTM < 1 CFU/100ml.
- Initiated
- 2018-04-20
- Root cause
- Device Design
- Terminated
- 2025-02-05
- Distribution
- MN, TX, WA, OR, IL, OH, SC, AK, PA, CA, WI, GA, NJ, FL, AZ, KY, MS, MO, LA, AR, MA, MI, MT, CO, TN, NE, NC, VA, NV, HI, PR, WY, AL, OK, IN, CT, ME, IA, SC, KS, NM, SD, DC, MD, ND,
- Recall number
- Z-1901-2018
- Recall status
- Terminated
- CDRH record ID
- 164360
- FDA classified
- 2018-05-18
- Recall event number
- 80072