Skip to content

Public FDA record

Z-1901-2021: Sapi Med sterile disposable colorectal anoscopes and proctoscopes. REF/Product Description: A.4083/THE B

Recall record Z-1901-2021 for Adler MicroMed, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1901-2021
Company in FDA record
Adler MicroMed, Inc.
FDA record creation date
Jun 21, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to issue with product sterility. Sterilization service provider had falsified records.

Source facts

Action
On 04/23/2021, the Recalling Firm communicated an URGENT: MEDICAL RECALL message via a telephone calls to customers and followed up email with a notification letter, informing them that the affected products that are labeled as sterile may not be sterile and the products could affect the sterile field and cause patient infections. Customers are instructed to: 1) Segregate and quarantine the affected products and wait for further instructions for disposition to follow, and 2) Complete and return the attached Acknowledgement and Receipt Form and indicate the quantities and location of the recalled products.
Initiated
2021-04-23
Root cause
Process control
Terminated
2022-10-04
Distribution
US distribution: FL, IL, MA, and PA.
Recall number
Z-1901-2021
Recall status
Terminated
CDRH record ID
187100
FDA record created
2021-06-21
Recall event number
87812

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗