Open original FDA record ↗Recall record
- FDA identifier
- Z-1901-2024
- Company in FDA record
- Smith & Nephew Inc
- FDA classification date
- May 24, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Packaging contained the incorrect size of femoral head from what was displayed on the label.
Source facts
- Action
- The recalling firm issued letters on 4/24/2024 informing the consignees of the issue, the patient impact, risks to health, and the actions to be taken by the users. The actions included to locate and quarantine affected devices immediately and if they have been further distributed to other organizations, the consignee is requested to inform their consignees of this field action and provide a copy of the letter. A Customer Response form is to be completed and emailed or FAX'd back to the national Smith + Nephew agency/distributor. The product is to be returned to the national Smith + Nephew agency/distributor.
- Initiated
- 2024-04-24
- Root cause
- Process control
- Distribution
- International distribution to the countries of Australia, Germany, Japan, Portugal, and United Kingdom.
- Recall number
- Z-1901-2024
- Recall status
- Open, Classified
- CDRH record ID
- 207564
- FDA classified
- 2024-05-24
- Recall event number
- 94555