Z-1901-2025: Access Testosterone assay, Catalog Number 33560

Recall record Z-1901-2025 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1901-2025
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Jun 2, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification letter dated 4/23/25 was sent to customers. ACTION: Discontinue using and discard all remaining Access Testosterone reagent packs from the lot listed above. Beckman Coulter recommends performing a retrospective review on patient results generated with the Access Testosterone lot listed above and assessing the need for retesting. Contact your local Beckman Coulter representative for replacement product requests. Beckman Coulter is no longer distributing the affected Access Testosterone reagent pack lot number and has corrected the manufacturing issue that led to cross-contamination. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: Electronically, if you received this communication via email. Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact our Customer Support Center: From our website: http://www.beckmancoulter.com If you need replacement product: o Complete the attached Replacement Order Form and email to askbeckman@beckman.com or fax to (866) 294-7850 OR o Call Client Services at (800) 526-3821
Initiated
2025-05-05
Root cause
Equipment maintenance
Distribution
US State: FL. UAE, Vietnam
Recall number
Z-1901-2025
Recall status
Open, Classified
CDRH record ID
213728
FDA classified
2025-06-02
Recall event number
96824

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗