Z-1902-2021: Ventec Life Systems VOCSN Multi-Function Ventilator Model/Version/Catalog Number: V+O+C+S+N+Pro/English/

Recall record Z-1902-2021 for Ventec Life Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1902-2021
Company in FDA record
Ventec Life Systems, Inc.
FDA classification date
Jun 21, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a component failure, ventilators may unexpectedly shut down or have a loss of blower function while in patient use. The shutdown potentially prevents the ventilator from providing ventilation therapy.

Source facts

Action
During the week of May 10, 2021, the firm sent an URGENT: MEDICAL DEVICE RECALL NOTIFICATION letter via FedEx or overnight mail to customers informing them that due to malfunctions of the Control Board there is the potential of unexpected shut downs and/or lost of ventilator blower function. As a result, the ventilators may fail to provide ventilation therapy to the patients. Customers are instructed to complete the following clinical action until a replacement unit can be provided: Follow your procedures to transfer the patient to an alternative source of ventilation, if available. If no other ventilation method is available, you may continue to use the affected product adhering to the guidelines below: 1. Use only when the patient is actively monitored to ensure timely response to a potential malfunction. 2. Operate the ventilator as defined in the Instructions for Use and promptly respond to all alarms activated by the ventilator. 3. Use an external SpO2 monitor and set the alarm thresholds appropriately if the patient is oxygen dependent. 4. Connect the ventilator to an external monitoring system, wherever possible. In addition, Customers are to acknowledge receipt of the Customer Notification Letter by: 1. Confirm receipt of this notification letter by returning the enclosed Customer Response Form to Ventec within 7 days. 2. Upon receipt of the Customer Response Form, Ventec will arrange shipment of replacement device(s) or arrange repair of existing device(s). 3. Return the affected product to Ventec per instructions and shipping label provided with replacement device(s). Contact Ventec Customer Support at 844-MYVOCSN (844-698-6276) to discuss alternative arrangements, if necessary. If customers are no longer the owner of any of these devices, they need to complete an ownership transfer form at VentecLife.com/tracking so that the Recalling Firm can follow up with the current owner of the impacted device(s). If customers have any question
Initiated
2021-05-10
Root cause
Device Design
Distribution
Distribution US nationwide, Hong Kong, and Japan
Recall number
Z-1902-2021
Recall status
Open, Classified
CDRH record ID
187222
FDA classified
2021-06-21
Recall event number
87838

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗