Open original FDA record ↗Recall record
- FDA identifier
- Z-1903-2012
- Company in FDA record
- Polymer Technology Systems, Inc.
- FDA classification date
- Jun 28, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
During a routine product investigation, a creatinine sample with a value less than 1.0 mg/dl was tested on a version 2.60 CardioCek PA meter and the meter provided a result of 0.08mg/dl, which is not the reportable measuring range for the CardioChek PA meter for the creatinine assay.
Source facts
- Action
- Polymer Technology Systems, Inc sent a URGENT FIELD CORRECTION letter dated May 15, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter instructed the customers to discontinue use and dispose of the product. PTS Inc. will provide appropriate replacement for any product that is properly identified on the attached Customer Acknowledgement and confirmation letter, If you have any questions please contact Customer Service toll-free at 1-877-870-5610, or +1-317-870-5610 ( outside the US )
- Initiated
- 2012-05-15
- Root cause
- Software design
- Terminated
- 2012-07-20
- Distribution
- Worldwide Distribution -- Nationwide Distribution including the states of CA and VA., and the countries of China, Poland, Ireland and France.
- Recall number
- Z-1903-2012
- Recall status
- Terminated
- CDRH record ID
- 109606
- FDA classified
- 2012-06-28
- Recall event number
- 61966