Skip to content

Public FDA record

Z-1903-2017: Dressing Rect L, P/N 02428 Product Usage: These products are sterile dressings used as an insulation bar

Recall record Z-1903-2017 for Breg Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1903-2017
Company in FDA record
Breg Inc
FDA classification date
May 2, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

They may not have been sealed prior to sterilization. Products with unsealed pouches will be non-sterile.

Source facts

Action
Breg sent an Urgent Medical Device Correction Letter to customers on 4/13/17. The letter identified the affected product, problem and actions to be taken. Customers were instructed to: Please take the following actions: 1) Determine if any potentially affected products are in your stock by comparing each product's Lot and Date to the Lot Numbers and Expiration Dates in the table above. Refer to Picture 2 for the location of these fields on the product packaging. 2) Once you have matched Lot and Date, inspect the product to identify any open (unsealed) pouches. Refer to Picture 1 below. 3) Remove any open pouches from your inventory and discard. 4) Complete the Response Form included as Attachment A, and email or fax to Breg as follows: a. E-mail: bregfieldcorrection@breg.com: Subject: Sterile Dressing Correction b. Fax: 800-FAX-BREG (329-2734)- Subject: Sterile Dressing Correction Customers with questions are instructed to contact Breg Customer Care at (800) 321-0607.
Initiated
2017-03-30
Root cause
Employee error
Terminated
2017-06-23
Distribution
US Nationwide Distribution
Recall number
Z-1903-2017
Recall status
Terminated
CDRH record ID
154790
FDA classified
2017-05-02
Recall event number
76994

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗