Skip to content

Public FDA record

Z-1903-2020: Tibial Tray Nonporous A Size 2; Model # PN 162-1200-A; Material: Ti6Al4V.

Recall record Z-1903-2020 for Ortho Development Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1903-2020
Company in FDA record
Ortho Development Corporation
FDA record creation date
May 6, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The locking mechanism of the tibial tray exhibited a manufacturing defect in which a uniform ridge was cut along the inner wall of the "T" and "Y" junctions, which may cause resistance when the Tibial Insert was being assembled.

Source facts

Action
The firm notified their US sales managers via teleconference on 07/19/2019. The sales managers coordinated distribution of the notification letter to the distributors and surgeons. All 12 US surgeons received both in-person verbal and written notices (hand delivered by sales reps) subsequent to the 07/19/2019 teleconference meeting. Recall actions for products sold in Japan were coordinated by the firm's parent company. In the firm's "Urgent Advisory Notice/Medical Device Recall Notice", dated 07/16/2019 includes the following instructions for end users: Contacting affected patients should be at the discretion of the surgeon to discuss the risks identified. The surgeon should weigh the risk of removing an implanted device against the risk of leaving it in place. Absence of clinical failure should be considered during the decision-making process. The appropriate action is left to the judgment of the surgeon and the preference of the patient. Complete and return the Customer Acknowledgement form. For any questions or concerns, call 801-553-9991 or contact via email at Ryan_Carpenter@orthodevelopment.com.
Initiated
2019-05-30
Root cause
Nonconforming Material/Component
Distribution
US: CA, FL, OH, PA, TX, and UT. OUS: Japan.
Recall number
Z-1903-2020
Recall status
Open, Classified
CDRH record ID
180188
FDA record created
2020-05-06
Recall event number
85157

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗