Open original FDA record ↗Recall record
- FDA identifier
- Z-1903-2025
- Company in FDA record
- Luminex Corporation
- FDA classification date
- Jun 3, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
It was determined that the consumable issue may result in No Call - INT CTL 1 and/or False Negative results on a small portion of the lot of test cartridges.
Source facts
- Action
- Diasorin issued an Urgent Field Safety Notice to its consignees on 04/16/2025 via email. The notice explained the issue, potential risk to health, and requested the following: "Actions to be taken by the Customer/User 1. Please examine your inventory for the affected products, cease all use, and isolate or quarantine them pending disposal. 2. If you currently have any of the identified affected barcode serial numbers from VERIGENE BC-GN Nucleic Acid Test Cartridge, Lot 022525021A, please contact Technical Support at support@luminexcorp.com to arrange for replacement and confirm the disposal of any remaining tests in your inventory per the package insert disposal guidelines. 3. Perform a retrospective review of results for VERIGENE Gram-Negative Blood Culture Nucleic Acid Test (BC-GN). If you have encountered a false negative result with a patient sample tested with the potentially impacted portion of Lot 022525021A, please contact Technical Support at support@luminexcorp.com. 5. If you are a distributor, provide a copy of this letter to any customers or organizations who may have received VERIGENE BC-GN Nucleic Acid Test Cartridge, Lot 022525021A. 6. Please complete the Acknowledgement & Receipt Form and return it to Technical Support. 7. For those with complaints, questions, or concerns please contact Technical Support at 1-877-785-2323 (U.S) or +1-512-381-4397 (Outside U.S.)."
- Initiated
- 2025-04-16
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution in the states of AL, CA, MN, TX, AZ, NJ, WA, SC, VA, GA, VT, CT, MD.
- Recall number
- Z-1903-2025
- Recall status
- Completed
- CDRH record ID
- 213425
- FDA classified
- 2025-06-03
- Recall event number
- 96701