Z-1903-2025: VERIGENE Gram Negative Blood Culture (BC-GN) Nucleic Acid Test, REF 20-006-021; for use with the sample-t

Recall record Z-1903-2025 for Luminex Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1903-2025
Company in FDA record
Luminex Corporation
FDA classification date
Jun 3, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was determined that the consumable issue may result in No Call - INT CTL 1 and/or False Negative results on a small portion of the lot of test cartridges.

Source facts

Action
Diasorin issued an Urgent Field Safety Notice to its consignees on 04/16/2025 via email. The notice explained the issue, potential risk to health, and requested the following: "Actions to be taken by the Customer/User 1. Please examine your inventory for the affected products, cease all use, and isolate or quarantine them pending disposal. 2. If you currently have any of the identified affected barcode serial numbers from VERIGENE BC-GN Nucleic Acid Test Cartridge, Lot 022525021A, please contact Technical Support at support@luminexcorp.com to arrange for replacement and confirm the disposal of any remaining tests in your inventory per the package insert disposal guidelines. 3. Perform a retrospective review of results for VERIGENE Gram-Negative Blood Culture Nucleic Acid Test (BC-GN). If you have encountered a false negative result with a patient sample tested with the potentially impacted portion of Lot 022525021A, please contact Technical Support at support@luminexcorp.com. 5. If you are a distributor, provide a copy of this letter to any customers or organizations who may have received VERIGENE BC-GN Nucleic Acid Test Cartridge, Lot 022525021A. 6. Please complete the Acknowledgement & Receipt Form and return it to Technical Support. 7. For those with complaints, questions, or concerns please contact Technical Support at 1-877-785-2323 (U.S) or +1-512-381-4397 (Outside U.S.)."
Initiated
2025-04-16
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of AL, CA, MN, TX, AZ, NJ, WA, SC, VA, GA, VT, CT, MD.
Recall number
Z-1903-2025
Recall status
Completed
CDRH record ID
213425
FDA classified
2025-06-03
Recall event number
96701

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗