Z-1904-2021: Philips Allura Xper System: Interventional Fluoroscopic X-Ray System Product No.: 722026, 722027, 722028,

Recall record Z-1904-2021 for Philips Medical Systems Nederlands. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1904-2021
Company in FDA record
Philips Medical Systems Nederlands
FDA record creation date
Jun 21, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Certain low-voltage DC power supplies (DCPS) used in these products may be subject to an increased probability of failure. Failure of a DCPS may result in the sudden loss of imaging functionality or mechanical movement, depending on what subsystems the DCPS is powering.

Source facts

Action
On April 23, 2021, Philips Healthcare issued an Urgent Medical Device Correction notice to customers via letter notifying them that Philips Healthcare has discovered through trend analysis an increase in the failure rate of certain low-voltage DC power supplies (DCPS) used in these products. Each system contains multiple DCPS, some of which may be subject to an increased probability of failure. Failure of a DCPS may result in the sudden loss of imaging functionality or mechanical movement, depending on what subsystems the DCPS is powering. HOW TO IDENTIFY AFFECTED PRODUCTS: For systems mentioned above, If 'Rev 8.2.16 or Rev 8.2.17 appears in the lower (black) area of the start-up screen, the system is affected. Philips will be notifying customers with affected systems. ACTION TO BE TAKEN BY CUSTOMER / USER: The likelihood of a system failure occurring is remote, and the users are recommended to follow their pre-established procedures for managing potential patient safety in the event that the system shuts down. ACTIONS PLANNED BY PHILIPS: Philips will replace the affected DCPS. This service will be provided free of charge for all affected systems. A Philips Healthcare service representative will contact customers with affected devices to arrange for the service. .
Initiated
2016-06-23
Root cause
Component design/selection
Terminated
2023-02-08
Distribution
Worldwide distribution. US states of: AL, AZ, FL, GA, IL, MA, MO, NJ, OR, and VA; Netherlands, United Kingdom, Poland, Spain, France, Czech Republic, India, Brazil, Singapore, Japan, Colombia, United Arab Emirates, China, Russian Federation, Oman, and Mexico.
Recall number
Z-1904-2021
Recall status
Terminated
CDRH record ID
187573
FDA record created
2021-06-21
Recall event number
87978

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗