Skip to content

Public FDA record

Z-1904-2026: OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2116.

Recall record Z-1904-2026 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1904-2026
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Apr 22, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.

Source facts

Action
Olympus notified consignees on about 03/25/2026 via letter titled "URGENT: MEDICAL DEVICE REMOVAL." Consignees were notified of the issue and hazard involved and were instructed to examine inventory and OER-ELITE devices for the affected product. The current version has black lock levers. The previous versions have gray lock levers or integrated lock levers, which are the same color as the rest of the connector. Consignees were instructed that if they have sufficient quantities of the current version (black) to support reprocessing needs, stop using the previous version and set aside for return. They may proceed with using the current version, only. If they have insufficient quantities of the current version (black) available, and do not have another method of endoscope reprocessing readily available, they may continue using the previous version(s) consistent with the Instructions for Use, which require users to inspect the connecting tube prior to use. Transition to the current version as soon as possible by placing an order through Customer Service at 1-800-848-9024, option 2. When sufficient quantities of the current version (black) are obtained to support reprocessing needs, stop using the previous version and only use the current version going forward. Then arrange for return the previous version of the connector. Consignees were also requested to confirm receipt of the letter through the Olympus web portal and forward the notification to other users who may have the affected products if product were further distributed.
Initiated
2026-03-25
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, China, Hong Kong, Japan, South Korea and Taiwan.
Recall number
Z-1904-2026
Recall status
Open, Classified
CDRH record ID
219040
FDA classified
2026-04-22
Recall event number
98544

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗