Z-1905-2020: Visualase Cooled Laser Applicator System Tubing Set kits Product numbers: VCLAS 9735559, 3mm tip/1.65mm

Recall record Z-1905-2020 for Medtronic Navigation, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1905-2020
Company in FDA record
Medtronic Navigation, Inc.
FDA record creation date
May 6, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm received complaints on the saline tubing of Visualase Cooled Laser Applicator System (VCLAS) kit, including incorrect Luer connector assembly, incorrect drip-chamber assembly, and/or excessive adhesive causing tubing occlusion, all of which could result in disruption or prevention of saline flow.

Source facts

Action
The firm issued its first customer notification and customer response form on 7/8/2019. On 7/16/2019, a second notification was issued to allow customers to identify recalled product by lot number. On 10/3/2019, the firm issued a third customer notification to include three additional affected lots. The firm requested customers examine inventory for affected lot numbers and quarantine for return to the firm and to complete the attached customer response form.
Initiated
2019-07-08
Root cause
Nonconforming Material/Component
Terminated
2024-10-16
Distribution
US: AL, AZ, AR, CA, CO, CT, DC, FL, GA, IL, IN, KY, LA, ME, MD, MA, MI, MN, MO, NE, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WV, WI OUS: France, Germany, Greece, Iceland, Israel, Spain, Sweden, Switzerland, UK
Recall number
Z-1905-2020
Recall status
Terminated
CDRH record ID
180672
FDA record created
2020-05-06
Recall event number
85354

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗