Open original FDA record ↗Recall record
- FDA identifier
- Z-1905-2020
- Company in FDA record
- Medtronic Navigation, Inc.
- FDA record creation date
- May 6, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm received complaints on the saline tubing of Visualase Cooled Laser Applicator System (VCLAS) kit, including incorrect Luer connector assembly, incorrect drip-chamber assembly, and/or excessive adhesive causing tubing occlusion, all of which could result in disruption or prevention of saline flow.
Source facts
- Action
- The firm issued its first customer notification and customer response form on 7/8/2019. On 7/16/2019, a second notification was issued to allow customers to identify recalled product by lot number. On 10/3/2019, the firm issued a third customer notification to include three additional affected lots. The firm requested customers examine inventory for affected lot numbers and quarantine for return to the firm and to complete the attached customer response form.
- Initiated
- 2019-07-08
- Root cause
- Nonconforming Material/Component
- Terminated
- 2024-10-16
- Distribution
- US: AL, AZ, AR, CA, CO, CT, DC, FL, GA, IL, IN, KY, LA, ME, MD, MA, MI, MN, MO, NE, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WV, WI OUS: France, Germany, Greece, Iceland, Israel, Spain, Sweden, Switzerland, UK
- Recall number
- Z-1905-2020
- Recall status
- Terminated
- CDRH record ID
- 180672
- FDA record created
- 2020-05-06
- Recall event number
- 85354