Z-1905-2025: MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-C

Recall record Z-1905-2025 for VIRTUAL INCISION CORPORATION. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1905-2025
Company in FDA record
VIRTUAL INCISION CORPORATION
FDA classification date
Jun 4, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the hand controllers and the movement of the Surgical Minibot arms.

Source facts

Action
A "MEDICAL DEVICE CORRECTION NOTIFICATION" letter dated 4/22/25 was sent to customers. Actions to be taken by the Customer/User: Do not use the device and transition scheduled procedures to another minimally invasive surgical technique i.e. laparoscopic or other robotic assisted device. Remove the device from service using your internal institutional procedures (e.g. Lock-Out, Tag-Out) for device maintenance until Virtual Incision technicians service the devices on hand. There is no anticipated impact to the conduct of the clinical study under the IDE. Please complete the attached Acknowledgement and Receipt form and return to Virtual Incision to indicate you received this letter. Customer Service, Clinical Support Staff, abby.cooper@virtualincision.com,913-608-3214. Quality Assurance, Sr Director of Quality and Manufacturing Engineering, Brandon.williams@virtualincision.com, 308-529-1987. Regulatory Affair, VP of Clinical and Regulatory Affairs, Florence.Beck@virtualincision.com, 352-275-726. Type of Action by the Company: Virtual Incision will be working with your site representatives to schedule on-site servicing of the device with updated software for the Surgeon Control Console within 30 days from receipt of this letter. Immediate corrections have been implemented through the analysis of the system data. Results indicate that implemented software updates have resolved the issue that could contribute to this deficiency in the future. If you have any question, call 513-500-4459 or email: mirasupport@virtualincision.com.
Initiated
2025-04-23
Root cause
Software design
Distribution
US Distribution to States: CA, FL, NE
Recall number
Z-1905-2025
Recall status
Completed
CDRH record ID
213740
FDA classified
2025-06-04
Recall event number
96831

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗