Z-1906-2008: Physio-Control LIFEPAK 20e defibrillator/monitor, sold internationally only, is indicated for the termin

Recall record Z-1906-2008 for Physio Control, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1906-2008
Company in FDA record
Physio Control, Inc.
FDA classification date
Sep 16, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The LIFEPAK 20e defibrillator / monitors with software versions 058 have an increase in likelihood for an incorrect Shock Advisory Algorithm (SAS) decision if the Auto Analyze setting in AED mode is On, leading to incorrectly render of shock or no shock decision. When Auto Analyze is set to On in AED mode the device initiates the SAS analysis immediately (no waiting period or warning prior to anal

Source facts

Action
On April 16, 2008, a letter was sent to the firms customers, via certified mail explaining the recall, giving product and software version information, request for customer to forward the notification to all sites and recommending the following: "Keep the defibrillator in service... Check the Auto Analyze Setup Option for each of your devices. If Auto Analyze is set to ON, change the setting to one of the other two choices: OFF or After First Shock... Follow the enclosed instructions to confirm or change the Auto Analyze Setup Option setting...If you update your defibrillator with new software, confirm the Auto Analyze Setup Option is set to OFF or After First Shock". Instructions for changing the auto analyze setup option for the LP 12 and LP20 were attached to the letter. There was no software issued in response to this action. Since LP20 customers may purchase a software program allowing them to update their defibrillator software, the letter recommends confirming Auto Analyze Setup Option is set to OFF or After First Shock if new software is installed . The default setting for the software update is Auto-Analyze ON. Contact Technical Support at 1-800-442-1142, option 5 - 6:00 a.m. to 4:00 p,m, (Pacific), Monday - Friday or visit www.Physio-Control-notices.com/sas if you have questions.
Initiated
2008-04-16
Root cause
Software design
Terminated
2011-03-29
Distribution
Product distributed nationwide and includes: 2 distributors, 157 direct accounts, 2,069 hospitals, 51 Department of Defense accounts, 58 miscellaneous government accounts, 3,295 first responders, 43 international accounts. International accounts are located in the following countries: ARGENTINA, AUSTRALIA, BARBADOS, BOLIVIA, BRAZIL, CANADA, CAYMAN ISLANDS, CHILE, COSTA RICA, DOMINICAN REPUBLIC, GERMANY, GUAM, GUATEMALA, GUYANA, HONG KONG, JAMAICA, JAPAN, MEXICO, NETHERLANDS ANTILLES, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, PANAMA, PARAGUAY, PERU, PUERTO RICO, TRINIDAD, UNITED STATES, URUGUAY, and VIRGIN ISLANDS.
Recall number
Z-1906-2008
Recall status
Terminated
CDRH record ID
70845
FDA classified
2008-09-16
Recall event number
47822

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗