Skip to content

Public FDA record

Z-1906-2011: Penumbra brand Penumbra Coil System, Penumbra Coil 400, Product is manufactured and distributed by Penumb

Recall record Z-1906-2011 for Penumbra Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1906-2011
Company in FDA record
Penumbra Inc.
FDA classification date
Apr 12, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The diameter of a component of the device is out of tolerance specification; the Delivery Pusher pull wire which secures the coil in place can prolapse and allow premature detachment of the coil.

Source facts

Action
Penumbra sent an Urgent Voluntary Field Removal Notice dated March 3, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately inspect all stock of Penumbra Coil 400 and identify product from the Lots. Remove product from these Lots from inventory and place it in quarantine for return to Penumbra Inc. (Do not use product from these Lots). Complete and return the attached product identification/ return form with in 5 business days via fax to +1 510-217-6414. Penumbra employees will contact customer directly to arrange return and replacement of product. For questions call Regulatory Affairs at 510-748-3224 or e-mail RAQA@prenumbrainc.com.
Initiated
2011-03-04
Root cause
Process change control
Terminated
2011-06-06
Distribution
Worldwide Distribution - USA (nationwide) and the countries in the European Union (Denmark, Germany, Italy, Netherlands, Slovakia, Slovenia, Spain, United Kingdom and Switzerland) and South Africa.
Recall number
Z-1906-2011
Recall status
Terminated
CDRH record ID
98588
FDA classified
2011-04-12
Recall event number
58175

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗