Open original FDA record ↗Recall record
- FDA identifier
- Z-1906-2011
- Company in FDA record
- Penumbra Inc.
- FDA classification date
- Apr 12, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The diameter of a component of the device is out of tolerance specification; the Delivery Pusher pull wire which secures the coil in place can prolapse and allow premature detachment of the coil.
Source facts
- Action
- Penumbra sent an Urgent Voluntary Field Removal Notice dated March 3, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately inspect all stock of Penumbra Coil 400 and identify product from the Lots. Remove product from these Lots from inventory and place it in quarantine for return to Penumbra Inc. (Do not use product from these Lots). Complete and return the attached product identification/ return form with in 5 business days via fax to +1 510-217-6414. Penumbra employees will contact customer directly to arrange return and replacement of product. For questions call Regulatory Affairs at 510-748-3224 or e-mail RAQA@prenumbrainc.com.
- Initiated
- 2011-03-04
- Root cause
- Process change control
- Terminated
- 2011-06-06
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries in the European Union (Denmark, Germany, Italy, Netherlands, Slovakia, Slovenia, Spain, United Kingdom and Switzerland) and South Africa.
- Recall number
- Z-1906-2011
- Recall status
- Terminated
- CDRH record ID
- 98588
- FDA classified
- 2011-04-12
- Recall event number
- 58175