Z-1906-2013: Nova Max Plus Monitor System (Kit w/10 Count Vials International) Catalog Number: 45507 Product Usage:

Recall record Z-1906-2013 for Nova Biomedical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1906-2013
Company in FDA record
Nova Biomedical Corporation
FDA classification date
Aug 9, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Reports of false abnormally high glucose readings from test strips

Source facts

Action
Nova Biomedical U.S.A issued an Urgent Medical Device Voluntary Recall Letter dated July 19, 2013 via registered/certified letters to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately discontinue use of the affected product; contact Nova Diabetes Care Customer Service for replacement strips; do not stop testing blood glucose while waiting for replacement strips; if affected strips must be used follow precautions provided in notification letter. For questions call Nova Diabetes Care Customer Service at 1-800-681-7390. .
Initiated
2013-07-15
Root cause
Material/Component Contamination
Terminated
2017-08-01
Distribution
Worldwide Distribution - USA Nationwide and the countries of Chile, England , Peru, Nicaragua
Recall number
Z-1906-2013
Recall status
Terminated
CDRH record ID
120660
FDA classified
2013-08-09
Recall event number
65617

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗