Open original FDA record ↗Recall record
- FDA identifier
- Z-1906-2020
- Company in FDA record
- Intuitive Surgical, Inc.
- FDA record creation date
- May 6, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The instrument drape may tear while draping the Patient Side Cart presenting a breach in the sterile barrier.
Source facts
- Action
- On 10/25/2018, the firm sent an "Urgent Medical Device Correction" Notification Letter to Customers via FedEx (to OUS customers - emailed and hand-delivered) are asked to take actions during draping to avoid snagging the drape during installation and to continue following the instructions in the System User Manual. Representatives will be available by phone to answer questions related to the device correction. Acknowledgement form included with the recall letter is requested to be completed and returned to ISI asking that appropriate personnel be informed about the device correction.
- Initiated
- 2018-10-25
- Root cause
- Device Design
- Terminated
- 2024-04-05
- Distribution
- Updated information provided on 03/18/2020: FL, OH, TN, NJ, PA, MD, NY, CA, MN, NV, OR, HI, VA, TX, AL, CA, IL, CN, and NC. OUS: APAC (Asia Pacific) - South Korea
- Recall number
- Z-1906-2020
- Recall status
- Terminated
- CDRH record ID
- 175509
- FDA record created
- 2020-05-06
- Recall event number
- 83503