Skip to content

Public FDA record

Z-1906-2023: Icelock 125 Ratchet, REF: L-125000; Icelock 600XM Ratchet, REF: L-621200; Icelock 621 Ratchet, REF: L-621

Recall record Z-1906-2023 for Ossur H / F. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1906-2023
Company in FDA record
Ossur H / F
FDA classification date
Jun 13, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A mechanical lock used to connect and release a locking liner into a prosthetic socket have an attachment pin and lock that can wear down faster than expected, leading to the potential for the pin to disengage from the lock, leading to loss of suspension of the external prosthetic system that replaces a missing lower limb, which can lead to falling and injury.

Source facts

Action
On 5/12/23, recall notices were sent to customers informing them of the following: a. For affected lots in use, replace both lock body and attachment pin as soon as possible for users that have the following risk factors: - Trouble with socket fit due to volume changes in the residual limb. - Had lock replacement without having new attachment pin installed at the same time. - Have non-Ossur attachment pin in use. - Complaint about loose fit of the socket or locking issues. b. For other users, replace the affected lots of lock body and attachment pin during the next planned consultation, such as liner replacement. If a user suggests there is an issue with the locking mechanism, the lock and attachment pin should be inspected for wear, regardless of lot number. Replace the lock and attachment pin if there are signs of wear; instructions for inspection of wear are attached. c. INFORM applicable personnel handling the products about the added warnings in the Instructions for Use. Firm recommends that you reach out to your patients as soon as possible to inform them about the potential risk of the pin disengaging from the lock, leading to loss of suspension. d. Return unused affected devices. e. Pass this notice to all who need to be aware within your organization or to any other organizations/customers who received affected product. f. Maintain the notice for awareness. g. Complete and return the acknowledgement form to recalls@ossur.com If you have questions or concerns about this notification, please contact customer service. Tel: 1-800-233-6263.
Initiated
2023-05-12
Root cause
Component change control
Distribution
US: SC, FL,NM, PA, VA, CA, TX, MT, UT, OH, NV, KS, MI, GA, IA, WA, TN, AZ, MN, NC, IN, MS, AL, NY, ID, DE, OK, MA, WI, OR, KY, MD, CO, WV, LA, IL, NJ, CT, MO, NE, WY, AR, SD, PR
Recall number
Z-1906-2023
Recall status
Open, Classified
CDRH record ID
199879
FDA classified
2023-06-13
Recall event number
92207

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.