Open original FDA record ↗Recall record
- FDA identifier
- Z-1907-2011
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Apr 8, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
When using the 8V3c transducer in color mode, the derated Ispta 3 (Spatial Peak-Temporal Average Intensity) may exceed the limit stated in the instructions for use.
Source facts
- Action
- A Customer Safety Advisory Notification was initiated on 2/4/2011 and sent via certified mail. The letter identified the affected product and described when the problem occurs along with the associated potential risk. The firm recommends that users not use the 8V3c until the software issue is fixed. The firm is working to release a software update and customers' local Customer Service Engineer will install new software on their system when it is available. Customers should direct their questions to their local service support person. The information in the letter should be shared with all personnel within the organization who need to be aware.
- Initiated
- 2011-02-04
- Root cause
- Software design
- Terminated
- 2011-10-04
- Distribution
- Worldwide Distribution.
- Recall number
- Z-1907-2011
- Recall status
- Terminated
- CDRH record ID
- 97958
- FDA classified
- 2011-04-08
- Recall event number
- 58006