Z-1907-2019: EliA RF IgM Well, REF 14-5600-01 Product Usage: Usage: EliA RF IgM is intended for the in vitro quantit

Recall record Z-1907-2019 for Phadia Ab. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1907-2019
Company in FDA record
Phadia Ab
FDA record creation date
Jul 1, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for reporting low assay results

Source facts

Action
On May 10, 2019, Corporate Thermo Fisher Scientific issued URGENT FIELD SAFETY NOTICES to customers by FedEx. Actions To Be Taken: Please review patient results obtained from lot 0085 and make note of the following: 1. If you have any of the above referenced material on hand, please discontinue use immediately and discard per your internal procedures. Notify all affected users in your facility to do the same. 2. All samples reported as positive (> 5 IU/ml) are positive, however the true concentration could be higher than reported. 3. The true concentration of a reported sample equal to 3.5 5 IU/ml could potentially be a positive sample. 4. Negative results between 2 IU/ml and 3.5 IU/ml are potentially false negative results. 5. Negative results < 2 IU/ml could be higher, but would still be reported as a negative result with a 60% higher concentration. 6. Complete and return the Customer Response Form. If you have any questions concerning this issue, please contact Phadia US, Inc. Technical Support at 1-800-346-4364, option #2.
Initiated
2019-05-10
Root cause
Nonconforming Material/Component
Terminated
2020-06-02
Distribution
US Nationwide n the states of NJ, NY, MI, WV, TX including PR
Recall number
Z-1907-2019
Recall status
Terminated
CDRH record ID
173270
FDA record created
2019-07-01
Recall event number
82979

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗