Z-1907-2023: Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Model Number 0998-00-0800-55, Part Number 0998-00-08

Recall record Z-1907-2023 for Datascope Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1907-2023
Company in FDA record
Datascope Corp.
FDA classification date
Jun 14, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

The digital IFUs provided with the Cardiosave IABP during the software update to version D.00 incorrectly annotated compliance to standard 60601-1-12:2014 within the updated IFU.

Source facts

Action
An URGENT FIELD SAFETY NOTICE MEDICAL DEVICE CORRECTION notification letter was sent to customer on 5/4/23. Actions to be taken by the customer: " Please locate the CARDIOSAVE Hybrid IABPs with the serial numbers listed in this notice. " Please ensure that all Cardiosave Hybrid Intra-Aortic Balloon Pump users at your facility are aware of this notice. " Please forward this information to all current and potential Cardiosave Hybrid IABPs users within your hospital / facility. " Please contact your local Getinge Service Representative to arrange to have the affected IFU USB drives replaced. " Please complete and sign the attached URGENT FIELD SAFETY NOTICE RESPONSE FORM (page 3) to acknowledge that you have received this notification. Return the completed form to Datascope Corp./Getinge by e-mailing a scanned copy to fieldaction@getinge.com Actions to be taken by Getinge: Getinge will provide your facility with an updated IFU USB drive that contains reference to the vibration testing standard, 60601-1-12:2014. The original USB drive that was provided during the D.00 software update can be destroyed locally. This voluntary correction notification only affects the products listed on page 1; no other products are affected by this voluntary correction. We apologize for any inconvenience this Medical Device Correction may cause. If you have any questions, please contact your Datascope Corp./Getinge representative.
Initiated
2023-05-04
Root cause
Under Investigation by firm
Distribution
International distribution in the country of Germany.
Recall number
Z-1907-2023
Recall status
Open, Classified
CDRH record ID
199910
FDA classified
2023-06-14
Recall event number
92214

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗