Skip to content

Public FDA record

Z-1910-2025: ChannelCheck Convenience Pack, Model Numbers UCC-222CP and UCC-222-ESP (Sample)

Recall record Z-1910-2025 for Healthmark Industries Co., Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1910-2025
Company in FDA record
Healthmark Industries Co., Inc.
FDA classification date
Jun 5, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries conducted its own investigation and concluded that the Pre-filled Water Syringe had been improperly used outside of the intended purpose.

Source facts

Action
An "URGENT PRODUCT RECALL" notification letter was sent to customers beginning 5/2/25. Recommended Mitigations/ User Recommendations Please discontinue use of the Pre-Filled Water Syringes containing non-sterile water, which can be identified by a red cap. Any unused product can be returned for credit or replacement. Customer are requested to take the following actions 1. Ensure all affected product in your facility s inventory is no longer in use and has been properly quarantined. 2. Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected products have been transferred. 3. Contact Customer Service via email at HMCS@hmark.com or 800-521-6224 to return your unused affected product or if you have any questions. 4. Complete and return the response form Getinge by e-mailing a copy to recallresponses.qrc@getinge.com.
Initiated
2025-05-02
Root cause
Under Investigation by firm
Distribution
US Nationwide and Internationally to countries of: Canada, Malaysia.
Recall number
Z-1910-2025
Recall status
Open, Classified
CDRH record ID
213735
FDA classified
2025-06-05
Recall event number
96823

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗