Z-1911-2008: Phillips Medical Systems, SureSigns VS3 Vital Signs Monitor model numbers 863071, 863072, 863073, 863074.

Recall record Z-1911-2008 for Philips Medical Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1911-2008
Company in FDA record
Philips Medical Systems
FDA classification date
Sep 16, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Some users of SureSigns VS3 Signs Monitor may have assumed that the SpO2 Non-Pulsatile and SpO2 No Sensor technical alarms were enabled when the SureSigns VS3 Vital Signs Monitor was not in Interval non-invasive blood pressure (NBP) mode. These alarms are only enabled when interval NBP mode is selected. If when monitoring SpO2, the clinician does not notice when the sensor is no longer attached, the patient is not being monitored through Spo2.This could result in delay or lack of needed therapy. The labeling provided with the monitor did not fully describe this feature.

Source facts

Action
All affected customers will receive the FSN (Field Safety Notice )informing them about the labeling correction. Included with the FSN will be an addendum to the instructions for use. customers are asked to read the addendum and insert a copy into each of their copies of the VS3 instructions for use. Customers will also receive a Medical Device Correction Confirmation Card that they will be asked to fill out and fax back to their local Philips representative. Contact your local Philips support organization at 1-800-722-9377 in the United States or 1-800-323-2280 in Canada.
Initiated
2008-04-30
Root cause
Labeling design
Terminated
2017-03-29
Distribution
Worldwide Distribution to United States, Canada and Foreign countries
Recall number
Z-1911-2008
Recall status
Terminated
CDRH record ID
70436
FDA classified
2008-09-16
Recall event number
47886

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗