Open original FDA record ↗Recall record
- FDA identifier
- Z-1912-2016
- Company in FDA record
- Synthes (USA) Products LLC
- FDA classification date
- Jun 6, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Devices did not pass the biological safety evaluation for cytotoxicity following exposure to test conditions. The high growth inhibition levels observed during testing could be attributed to corrosion of the device at solder points. This could potentially be reproduced during use and reprocessing.
Source facts
- Action
- Depuy Synthes sent an Urgent Notice: Medical Device Recall letter on 05/02/2016 to all affected consignees. Consignees were instructed to do the following: review your inventory immediately to see if any of the affected products were being stored, and call DePuy Synthes at 1-800-479-6329 to obtain a Return Authorization (RA) Number and replacement order, to complete the Verification Section (page 3) and return with the product to Credit/Returns, DePuy Synthes, 1101 Synthes Avenue, Monument, CO 80132.
- Initiated
- 2016-05-02
- Root cause
- Process design
- Terminated
- 2016-12-13
- Distribution
- US nationwide distribution.
- Recall number
- Z-1912-2016
- Recall status
- Terminated
- CDRH record ID
- 146229
- FDA classified
- 2016-06-06
- Recall event number
- 74080