Open original FDA record ↗Recall record
- FDA identifier
- Z-1912-2019
- Company in FDA record
- Teleflex Medical
- FDA classification date
- Jul 15, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Reported complaints of cracks being observed, prior to use, on swivel wye adaptors of the Neonatal ConchaSmart Breathing Circuit.
Source facts
- Action
- The firm initiated the recall by email on May 10, 2019. The recall notice requested the return of the product. The letter identified the affected product, problem and actions to be taken. Customers were instructed to 1. If you have affected stock in inventory, immediately discontinue use and quarantine the products. 2. Please complete the enclosed Acknowledgement Form as soon as possible indicating whether you do or do not have stock and fax it to 1-855-419-8507 or email to recalls@teleflex.com. A customer service representative will contact you with a Return Goods Authorization (RGA) number and provide instructions for the return of products. For questions contact Customer Service at 1-866-396-2111.
- Initiated
- 2019-05-10
- Root cause
- Process control
- Terminated
- 2024-04-22
- Distribution
- US Nationwide Distribution in the states of AL, TX
- Recall number
- Z-1912-2019
- Recall status
- Terminated
- CDRH record ID
- 173297
- FDA classified
- 2019-07-15
- Recall event number
- 82994