Skip to content

Public FDA record

Z-1914-2019: Atellica CH Creatine Kinase (CK_L) Reagent, Product Code 11097640, UDI Number: 00630414006642 - Product U

Recall record Z-1914-2019 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1914-2019
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA record creation date
Jul 2, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A steady upward trend in blank (u) absorbance was observed

Source facts

Action
Siemens Healthineers issued Urgent Medical Device Recall notices to domestic customers and Urgent Field Safety Notices to international customers dated May 8, 2019. Actions to be Taken by the Customer 1) Discontinue use of and discard the kit lots listed on the notice. 2) Review your inventory of these products to determine your laboratorys replacement needs and provide information to Siemens for reporting to the authorities. 3) Complete and return the Field Correction Effectiveness Check letter within 30 days. 4) Please review the letter with your Medical Director. 5) If you have received any complaints of illness or adverse events associated with the products listed in the notice, immediately contact your local Siemens Customer Care Center or your local Siemens technical support representative. 6) Please retain the letter with your laboratory records and forward the letter to those who may have received this product. If you have any questions, please contact your Siemens Customer Care Center or your local Siemens technical support representative.
Initiated
2019-05-03
Root cause
Nonconforming Material/Component
Terminated
2021-01-19
Distribution
Worldwide distribution.
Recall number
Z-1914-2019
Recall status
Terminated
CDRH record ID
173278
FDA record created
2019-07-02
Recall event number
82980

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗