Z-1915-2025: Newport HT70 Ventilator, REF: HT70M-JP-NA, HT70M-CN-NA, HT70M-ES-EU, HT70M-PT-BR, HT70M-SY-EU, HT70M-WW-E

Recall record Z-1915-2025 for Covidien. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1915-2025
Company in FDA record
Covidien
FDA classification date
Jun 9, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Ventilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation, or the shutdown alert alarm fails to alarm effectively during shut down, which may result in respiratory failure, hypoventilation, low oxygen saturation, hypoxia, treatment delay.

Source facts

Action
On 05/21/2025 recall notices were mailed and/or emailed to customers who were asked to do the following: Healthcare providers with patients currently ventilated with affected ventilators should assess overall risk and appropriate timing for device replacement. - Patient connected to a ventilator requires the constant attention of trained caregivers to the patient's condition. - Have an alternate power source and means of ventilation available when the ventilator is in use. - Use appropriate monitors to ensure sufficient oxygenation and ventilation when ventilator is in use. - If the patient is not responding to ventilation appropriately, the patient should be taken off the ventilator immediately and connected to an alternate method of ventilation. Contact the patient's health care provider/physician. - Follow Patient Set-up Procedure described in Operator's Manuals. Medtronic is not correcting these issues on affected ventilators or service parts and will no longer service affected ventilators identified in this notification. -If affected ventilator or service part is currently in use, remove and replace it with an alternate means of ventilation. After completing the transition, quarantine all affected ventilators and service parts and dispose. Patients are asked to return affected devices to their healthcare provider. Complete/return the Acknowledgement Form via email to rs.gmbmitgfca@medtronic.com. Questions or concerns contact firm at 1-800-255-6774 Distribute notice to all relevant personnel within your organization and any organizations where ventilators or service parts may have been transferred or distributed. Provide the enclosed patient letter to any of your patients using this ventilator in the home care setting to inform them of this recall. On 8/11/2026, recall notices listing additional affected devices were mailed and/or emailed to customers.
Initiated
2025-05-21
Root cause
Component design/selection
Distribution
Worldwide - US Nationwide distribution including in the states of MA, CT, KS, OH, PR, NH, GA, WI, MS, VA, CO, TX, MN, AL, IA, NY, UT, NM, KY, RI, SD, PA, IA, CA, MO, HI, MI, IL, MT, NJ, TN, SC, FL, IN, AZ, ID, LA, WA, NC, MD, WY, NE, NV, ND, ME, AK, DE, OK, OR and the countries of Japan, Canada, Singapore, Lithuania, Kazakhstan, Slovenia, Colombia, China, Qatar, United Arab Emirates, Greece, Saudi Arabia, Nigeria, Romania, Netherlands, Lebanon, Oman, Jordan, Malawi, India, Croatia, Portugal, Turkey, Brazil, Pakistan, Trinidad and Tobago, Slovakia, Chile, Egypt, Switzerland, South Africa, Thailand, Bangladesh, Myanmar, Peru, Uruguay, Hungary, Argentina, Iran, Islamic Republic Of, Costa Rica, Vietnam, Australia, Guatemala, Mexico, Guyana, Jamaica, Dominican Republic, Bahamas, Bosnia and Herzegovina, South Korea, Tunisia, Ukraine, Uzbekistan, Indonesia, Burkina Faso, Italy, Panama, Sri Lanka, Malaysia, Namibia, Swaziland, Spain, Philippines, Bhutan, Mauritius, Poland, France, Cambodia, Czechia, Israel, Norway, Russian Federation, United Kingdom, Canary Islands, Kenya, Czech Republic, Croatia , Estonia, Maldova, Palestine.
Recall number
Z-1915-2025
Recall status
Open, Classified
CDRH record ID
213582
FDA classified
2025-06-09
Recall event number
96774

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗