Open original FDA record ↗Recall record
- FDA identifier
- Z-1916-2009
- Company in FDA record
- Stryker Howmedica Osteonics Corp.
- FDA classification date
- Aug 26, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Triathlon Cutting Blocks did not assemble to the Triathlon Impactor/Extractor Handle.
Source facts
- Action
- Important Product Correction letters were sent to all branches/agencies on February 12, 2008 by Federal Express. The letter asked consignees to examine their inventory and hospital locations to identify the affected product. Customers were asked to perform a functional inspection by assembling the Triathlon Impactor/Extractor Handle to all of their MIS Cutting Blocks. If the cutting block does not assemble, customers are to list the affected lot codes on the Attached Product Accountability Form and fax a copy to (201) 831-6069. The letter states that product should not be returned until further instructions are received. Questions should be directed to (201) 831-5825.
- Initiated
- 2007-02-12
- Root cause
- Process control
- Terminated
- 2009-10-15
- Distribution
- Worldwide Distribution
- Recall number
- Z-1916-2009
- Recall status
- Terminated
- CDRH record ID
- 83571
- FDA classified
- 2009-08-26
- Recall event number
- 52641