Z-1916-2018: Synthes Torque Limiting Handle, Part Number 03.231.013 This device is used in systems indicated for frac

Recall record Z-1916-2018 for Synthes (USA) Products LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1916-2018
Company in FDA record
Synthes (USA) Products LLC
FDA record creation date
May 23, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The adapter may demonstrate a degradation of function as the final products were assembled with ball bearings of high carbon content that is likely to rust prematurely.

Source facts

Action
An Urgent Medical Device Recall Notification letters dated May 4, 2018, were distributed to customers. The letter requests that customers take the following actions: Potential Patient Impact: The subject, Torque Limiting Handle, may demonstrate a degradation of function with the adapter collar sticking, and/or not retaining the mating adapter male shaft of the driver component. In addition, the ball bearings may corrode over time causing a failure in the torque limiting feature. The following potential scenarios could occur: " Malunion/nonunion and device loosening if there is a degradation of function and too much or too little torque is applied. " Adverse tissue reaction if there is corrosion and any debris enters an open incision. " Surgical delay due to the time required to investigate and decide next steps if the instrument fails during surgery. If you have no product to be returned: " Return the Verification Section (page 3 of this letter) to DePuy Synthes by: o Fax: 866-853-1969 or o Scan/email: Synthes7370@stericycle.com NOTE: If the Verification Section is completed on behalf of more than one facility and/or individual, please clearly indicate the name and address of the facility and/or individual in page number three (3) of this notification. 3. Forward this notice to anyone in your facility that needs to be informed. 4. If any of the subject product has been forwarded to another facility, contact that facility to arrange return. 5. Keep a copy of this notice visibly posted for awareness until all product subject to this recall (removal) have been returned. While processing your returns, please maintain a copy of this notice with the product subject to this recall (removal) and keep a copy for your records. For further questions, please call (610) 719-5450.
Initiated
2018-05-04
Root cause
Component design/selection
Terminated
2019-10-21
Distribution
USA (nationwide) Distribution to the states of : AK, CA, FL, ID, IN, KS, KY, LA, MD, MI, MT, NC, NJ, NY, OH, OK, PA, SD, TN, TX, UT, WI, and WV.
Recall number
Z-1916-2018
Recall status
Terminated
CDRH record ID
164278
FDA record created
2018-05-23
Recall event number
80048

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗