Open original FDA record ↗Recall record
- FDA identifier
- Z-1918-2015
- Company in FDA record
- New Star Lasers, Inc.
- FDA classification date
- Jun 27, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During internal testing a situation was discovered where when changing the settings, a new setting could not be accepted and potentially result in a discrepancy in the energy delivered (higher or lower) from the value displayed on the Graphical User Interface display.
Source facts
- Action
- Medical Device Correction letters were sent to all affected consignees by Federal Express on May 13, 2015. The letter identified the affected product as well as the issue and risk involved. The software will be upgraded to address the issue; and, the firm will contact customers to coordinate installation of the upgraded software. Customers may continue to use their system safely by taking the precautions provided in the letter. Customers are to complete the attached response form. Questions should be directed to Nina Davis at 916-677-1909.
- Initiated
- 2015-05-13
- Root cause
- Software design (manufacturing process)
- Terminated
- 2016-10-28
- Distribution
- Worldwide Distribution -- US, including the states of CA, IL, AZ, FL, OH, MA, MI, and AL; and the country of the Netherlands.
- Recall number
- Z-1918-2015
- Recall status
- Terminated
- CDRH record ID
- 137356
- FDA classified
- 2015-06-27
- Recall event number
- 71346