Skip to content

Public FDA record

Z-1918-2018: Inspire 6 Integrated Phisio OXY w/ HVR Reservoir, Sterile EO, Catalog # 050713 The intended uses for the

Recall record Z-1918-2018 for LivaNova USA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1918-2018
Company in FDA record
LivaNova USA
FDA record creation date
May 23, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Some integrated sterile INSPIRE family products contain an incorrect expiration date printed in the Unique Device Identification (UDI) bar code and readable code just beneath the barcode. Specifically, the expiration date is reported as DDMMYY instead of YYMMDD. The expiration date in the applicable field on the package labels are correct as printed.

Source facts

Action
On March 13, 2018, notification letters were emailed. Customers were instructed to return response forms, which indicated the following: amount of unused product, and whether they would contact the firm to relabel effected products or if they preferred to be trained on the relabeling procedure. If product was further distributed, customers were instructed to send the notification letter to these additional customers within 2 business days. No action is required for product already used since it was consumed prior to the true expiration date. For further questions, please call (281) 228-7260.
Initiated
2018-03-13
Root cause
Incorrect or no expiration date
Distribution
US Distribution to the states of : NJ, CA, MN, MO, MA
Recall number
Z-1918-2018
Recall status
Terminated
CDRH record ID
163774
FDA record created
2018-05-23
Recall event number
79863

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗