Skip to content

Public FDA record

Z-1918-2026: Rover Mobile X-ray System Model/Catalog Number: MXU-RV35 and MXU-RV71

Recall record Z-1918-2026 for Micro-X Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1918-2026
Company in FDA record
Micro-X Ltd.
FDA classification date
Apr 22, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

X-Ray generator may malfunction resulting in the X-Ray being inoperable.

Source facts

Action
On March 27, 2026, Micro-X issued a Urgent: Medical Device Correction Notification, via E-Mail. Micro-X ask consignees to take the following actions: " Contact your local service representative to arrange replacement of the PCBA with the updated version at the next scheduled visit or Preventative Maintenance. In the interim, customers may continue to use the device. " To minimize the likelihood of system failure, the batteries should not be permitted to fully discharge. Any system errors or abnormal behaviours must be reported to your authorised service provider without delay. " If the product has been further distributed to other facilities, please ensure this notice is immediately shared with them. " Please complete and email the Customer Response Form in Appendix 1 to the Micro-X LTD in Australia: djerkovic@micro-x.com and copy support@micro-x.com Completing this form confirms receipt of this notice, understanding of the issue, and the necessary actions to be taken.
Initiated
2026-03-27
Root cause
Component design/selection
Distribution
US Nationwide distribution in the states and territories of WA, NJ, CA, Puerto Rico.
Recall number
Z-1918-2026
Recall status
Open, Classified
CDRH record ID
219326
FDA classified
2026-04-22
Recall event number
98644

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗