Open original FDA record ↗Recall record
- FDA identifier
- Z-1919-2018
- Company in FDA record
- LivaNova USA
- FDA record creation date
- May 23, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Some integrated sterile INSPIRE family products contain an incorrect expiration date printed in the Unique Device Identification (UDI) bar code and readable code just beneath the barcode. Specifically, the expiration date is reported as DDMMYY instead of YYMMDD. The expiration date in the applicable field on the package labels are correct as printed.
Source facts
- Action
- On March 13, 2018, notification letters were emailed. Customers were instructed to return response forms, which indicated the following: amount of unused product, and whether they would contact the firm to relabel effected products or if they preferred to be trained on the relabeling procedure. If product was further distributed, customers were instructed to send the notification letter to these additional customers within 2 business days. No action is required for product already used since it was consumed prior to the true expiration date. For further questions, please call (281) 228-7260.
- Initiated
- 2018-03-13
- Root cause
- Incorrect or no expiration date
- Distribution
- US Distribution to the states of : NJ, CA, MN, MO, MA
- Recall number
- Z-1919-2018
- Recall status
- Terminated
- CDRH record ID
- 163800
- FDA record created
- 2018-05-23
- Recall event number
- 79863