Z-1919-2020: REF: Z3239, 126" 10 Drop Primary Set w/3 MicroClave, 2 Gang 4-Way Stopcocks, Rotating Luer, 1 Ext, Steril

Recall record Z-1919-2020 for ICU Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1919-2020
Company in FDA record
ICU Medical, Inc.
FDA record creation date
May 6, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop cannula could result in inaccurate delivery of the intended infusion volume (under-delivery). This could result in under-delivery of critical medication which may lead to serious adverse health consequences.

Source facts

Action
On October 4, 2019, a "Urgent: Medical Device Recall" Letter was sent to all affected customers via US Mail. In addition to providing customers information on the recalled products, the letter asked customers to do the following: 1) Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the attached response form. Return the completed response form to the fax number or e-mail address on the form, even if you do not have the affected product. 3) Return affected product using the return label provided with this letter. Contact Stericycle at 1-866-367-4703 (Monday - Friday, 8:00 am-5:00 pm ET) if you have not received a return label or require additional labels for returning the affected product. The return labels are for single use only. Please do not reproduce. Please visit http://expertezlabel.com to request additional labels for returning affected product. To ensure proper and timely credit, follow the instructions on the return label for returning product. Upon receipt of the completed response form and return of the affected product, ICU Medical will credit you for any product returned. You will only receive credit for product that you return. NOTE: Credits for product purchased through distributor will be credited by the distributor. 4) If you have distributed the product further, immediately notify your accounts that received the product identified in the Affected Product / Table 1 sections of this notification and ask them to contact Stericycle at 1-866-367-4703 (M-F, 8am-5pm ET) to obtain a response form. 5) Replacement product is available. Please contact customer service using the information provided below. - Global Complaint Management - 1-844-654-7780 or ProductComplaintsPP@icumed.com, To report adverse events or product complaints - C
Initiated
2019-10-04
Root cause
Mixed-up of materials/components
Terminated
2021-12-28
Distribution
US:AK,AL,AR,AZ,CA,CO,CT,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,SD,TN,TX,UT,VA,WA,WI,WY OUS: Canada *** "REF" numbers AH7117 and B30211 were not marketed or distributed within the US***
Recall number
Z-1919-2020
Recall status
Terminated
CDRH record ID
180022
FDA record created
2020-05-06
Recall event number
85053

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗