Skip to content

Public FDA record

Z-1919-2025: Kendall SCD Blended Comfort Sleeves Knee Length Small - For use with Kendall SCD SmartFlow SKU 84021B K

Recall record Z-1919-2025 for Cardinal Health 200, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1919-2025
Company in FDA record
Cardinal Health 200, LLC
FDA classification date
Jun 6, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Incorrect software logic may cause frequent E1 errors (system high pressure). Issue may cause delay in treatment or therapy.

Source facts

Action
On May 13, 2025, the firm began to notify customers via Urgent Medical Device Product Correction letters. Customers were informed of the issue, which exclusively impacts use of the Comfort Small Knee Sleeve. Other sleeve sizes are unaffected and operate as intended with the SmartFlow Controller. The product remains operational if the Comfort Small Knee Sleeve and SmartFlow Controller configuration is not utilized during therapy. A software update on the SmartFlow Controller is necessary before using the SmartFlow controller with the Comfort Small Knee sleeve configuration. Customers should discontinue use of the Comfort Small Knee Sleeve with the SmartFlow Controller and utilize a suitable alternative such as SM thigh or foot cuff. Customer service representatives will coordinate with customers to facilitate a software update which will allow the controllers to be utilized with the small knee sleeves without causing the error.
Initiated
2025-05-13
Root cause
Software change control
Distribution
Domestic US distribution to AK, CA, FL, MA, TX. International distribution to Japan.
Recall number
Z-1919-2025
Recall status
Completed
CDRH record ID
213780
FDA classified
2025-06-06
Recall event number
96849

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗