Skip to content

Public FDA record

Z-1920-2024: MEGADYNE MEGA SOFT Pediatric Patient Return Electrode. Reusable Patient Return Electrode for patients wei

Recall record Z-1920-2024 for Megadyne Medical Products, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1920-2024
Company in FDA record
Megadyne Medical Products, Inc.
FDA classification date
Jun 6, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Reports of patient burns.

Source facts

Action
Consignees were mailed an URGENT: MEDICAL DEVICE RECALL (REMOVAL) notification, dated 5/8/24. The notification instructs consignees to examine their inventory to identify affected devices, quarantine any on-hand devices subject to the recall with a copy of the recall notice, and communicate the recall notification to all those in consignee facilities that should be informed and those facilities to which product was further distributed. Consignees are asked to return a completed Business Reply Form by email to ethicon4284@sedgwick.com or by fax at 800-978-9467. All devices subject to this recall are to be returned immediately; for credit reimbursement customers must return product no later than 8/31/2024. To return product, consignees are to include a photocopy of the completed Business Reply Form in the box with returned devices and affix the pre-paid authorized shipping label on product. Extra shipping labels may be obtained by calling 888-912-7090. Consignees with any questions can contact Ethicon Resource Department at 1-877-384-4266 from 8:00 AM to 5:00 PM EST from Monday through Friday.
Initiated
2024-05-08
Root cause
Device Design
Distribution
Worldwide distribution. US nationwide including Puerto Rico; Australia, Austria, Bahrain, Belgium, Canada, Chile, Czechia, France, Germany, Greece, Hong Kong, India, Ireland, Italy, Japan, the Netherlands, Norway, Poland, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Switzerland, Thailand, United Arab Emirates, and United Kingdom.
Recall number
Z-1920-2024
Recall status
Open, Classified
CDRH record ID
207516
FDA classified
2024-06-06
Recall event number
94544

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗