Three lots of Anthology HO Porous size 7 stems were manufactured with an undersized taper length. The device taper length was incorrectly manufactured as 0.393mm opposed to 0.442 mm.
Source facts
Action
The firm initiated the recall by letter on 05/02/2018. Consignees are asked to inspect inventory, locate and quarantine affected, unused devices. Affected product will be returned to Smith & Nephew. All consignees are asked to complete the Response Form, even if they do not have product to return. Distributors were directed to notify their customers of the field action and ensure that these actions are carried out. For further questions, please call (978) 749-1440
Initiated
2018-05-02
Root cause
Process control
Terminated
2020-09-01
Distribution
US nationwide distribution.
Recall number
Z-1921-2018
Recall status
Terminated
CDRH record ID
164289
FDA classified
2018-05-09
Recall event number
80050