Open original FDA record ↗Recall record
- FDA identifier
- Z-1921-2021
- Company in FDA record
- Exact Medical Manufacturing, Inc.
- FDA record creation date
- Jun 21, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
No 510k for the product to be used in a natural or surgical opening to the body
Source facts
- Action
- Exact Medical issued Correction Notification via email on 4/29/21. Letter states: Exact Medical Manufacturing Ultrasonic Transducer Cover can be used to minimize contamination between patient and ultrasound probe during ultrasound scanning procedures for external intact skin. This may help with easier cleaning and disinfection of the probe. This product is not to be used in a natural or surgical opening to the body. The Exact Medical Manufacturing Ultrasonic Transducer Cover are also sold as bulk non-sterile, single use items, to repackers/relabelers establishments for further packaging and ethylene oxide sterilization. These are the boundaries and limitations of use for existing Probe/Transducer cover products in any existing stock from Exact Medical Manufacturing, Inc Further instructions will be given at a future date and we encourage you to reach out with any questions or concerns regarding this issue.
- Initiated
- 2021-04-29
- Root cause
- No Marketing Application
- Terminated
- 2022-05-26
- Distribution
- US distribution to states of FL, MT, NY, OH, GA, and Puerto Rico
- Recall number
- Z-1921-2021
- Recall status
- Terminated
- CDRH record ID
- 187555
- FDA record created
- 2021-06-21
- Recall event number
- 87958