Skip to content

Public FDA record

Z-1922-2019: SOMATOM go.Top, Model No. 11061648 with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product U

Recall record Z-1922-2019 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1922-2019
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Jul 2, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The potential sporadic performance problems may cause scanning workflow interruptions and unexpected user notifications resulting in diagnostic delay or need for patient rescan.

Source facts

Action
On June 4, 2019, the firm notified customers of the recall via Urgent Medical Device Correction letters. Customers were informed that the firm has developed software update CT VA20A-SP2A to ensure uninterrupted scanning workflows and to reduce the number of user notifications. This update will also improve some areas during the start-up and shut-down processes, as well as providing other improvements related to the system, examination and reconstruction processes, and dual energy and intervention . The corrective action will be provided free of charge and will be distributed via Update Instruction CT026/19P. This update will be performed remotely. If you have any questions, please contact our service organization at 1- 800-888-7436. Customers were asked to promptly notify and instruct all staff at their organizations who need to be aware of this notice. In addition, if you have further distributed this product, please identify your customers and notify them at once of this product advisory notice. If the device has been sold and is therefore no longer in your possession, please forward this notice to the new owner.
Initiated
2019-06-04
Root cause
Software design
Terminated
2023-03-29
Distribution
US Nationwide distribution in the states of AL CA CT FL GA IL IN KS KY LA ME MI MO MS NC ND NE NH NY OH OK SD TN TX VA and WI
Recall number
Z-1922-2019
Recall status
Terminated
CDRH record ID
173752
FDA record created
2019-07-02
Recall event number
83112

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗