Skip to content

Public FDA record

Z-1923-2016: 4 ml FX Sodium Fluoride/Potassium Oxalate tube, Ref 454297 VACUETTE Glycolytic Inhibitor Tubes contain an

Recall record Z-1923-2016 for Greiner Bio-One North America, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1923-2016
Company in FDA record
Greiner Bio-One North America, Inc.
FDA record creation date
Jun 7, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The 4 ml FX Sodium Fluoride/Potassium Oxalate tube may have no or low additive (sodium fluoride and potassium oxalate), which can effect analytical results for glucose and lactate.

Source facts

Action
Greiner Bio-One North America, Inc. sent an Urgent Product Recall letter dated April 26, 2016, to all affected consignees. The letter requested that they stop distribution of the product and to provide Greiner with a list of their sub-accounts. The letter included a Product Disposition/Site Confirmation form which is to be returned to Greiner Bio-one. Consignees with questions were instructed to call 800-515-8112. For questions regarding this recall call 704-261-7823.
Initiated
2016-04-26
Root cause
Process control
Terminated
2016-08-22
Distribution
Nationwide Distribution to CA, FL, IL, KY, NY, PA, VA and Puerto Rico
Recall number
Z-1923-2016
Recall status
Terminated
CDRH record ID
145883
FDA record created
2016-06-07
Recall event number
74002

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗