Z-1923-2019: Philips HeartStart MRx devices in use with the M3539A AC Power Module: M3535A, M3536A, M3536M, M3536MC, M

Recall record Z-1923-2019 for Philips North America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1923-2019
Company in FDA record
Philips North America, LLC
FDA classification date
Jul 5, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The AC power module may fail at a higher than expected rate. If a fully charged battery is not properly installed, the HeartStart MRx may lose all power and fail to operate. Interruption of monitoring and failure or delay in the delivery of a shock or pacing can occur. A faulty AC Power Module may also fail to charge the battery.

Source facts

Action
On June 7, 2019, the firm mailed Urgent Medical Device Correction letters to customers, informing them that the AC Power Module may fail, which may lead to an interruption of monitoring or a delay in the delivery of a shock or pacing therapy. A failed AC Power Module also will not charge the battery. Customers were asked to take the following actions: 1. Per the Instructions for Use (IFU), the HeartStart MRx Monitor/Defibrillator can be powered by either a Lithium Ion battery or AC mains power (using the M3539A AC Power Module). However, a charged battery should always be installed in the device, whether or not AC mains power is available at the point of care. It is also important to keep your batteries charged. 2. Verify that the AC Power Module is functioning properly: a. With the device is plugged into AC mains power, assure that the External Power Indicator on the MRx display is green, as shown in Fig. 1. b. Verify that the battery is charging or charged by checking the Battery Charge Level Indicators shown in Fig. 2 below, as a faulty AC Power Module may fail to charge the battery. 3. Verify your MRx does not show a red X nor is it emitting a periodic audio chirp as this could be an indicator that there is no battery in the unit or that the AC Power Module may have failed, and the battery charge has dropped below 20%. 4. If you identify a failed AC Power Module using the criteria above, please contact Philips for a replacement. 5. To acknowledge receipt of this notification, please complete and fax the Customer Reply Form to: +1.877.499.7223 or email to recall.response@philips.com. The firm will provide a replacement for failed AC power modules to the customer at no charge. If you need further information or support concerning this notification, please contact your local Philips representative or call the firm at 1-800-722-9377.
Initiated
2019-06-07
Root cause
Process control
Terminated
2020-05-08
Distribution
Nationwide domestic distribution. Worldwide distribution.
Recall number
Z-1923-2019
Recall status
Terminated
CDRH record ID
172626
FDA classified
2019-07-05
Recall event number
82750

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗