Skip to content

Public FDA record

Z-1923-2026: Oxoid Agglutinating Sera, Salmonella 9-0 R30957301

Recall record Z-1923-2026 for Remel Europe Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1923-2026
Company in FDA record
Remel Europe Ltd.
FDA classification date
Apr 22, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Affected lots are not performing to IFU criteria showing cross reactions with 2-0 within 60 seconds. Cross-reactivity may result in the misidentification of a Salmonella 9-O serotype as a 2-O serotype.

Source facts

Action
Firm began notifying consignees on March 25, 2026 via letters titled "Urgent: Medical Device Recall." Laboratory professionals are advised to review results and destroy any remaining inventory of affected lots. Affected lots should not be used for any clinical laboratory testing. Customers should complete and return the included acknowledgement form for replacement/credit to be processed. Notice should be passed on to all who need to be aware. Firm is investigating the root cause and will implement corrective and preventive actions to prevent future recurrence.
Initiated
2026-03-25
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution to the states of AL, CA, FL, IA, IL, IN, MI, MN, MO, NC, NE, OH, SD, TX, WA.
Recall number
Z-1923-2026
Recall status
Open, Classified
CDRH record ID
219344
FDA classified
2026-04-22
Recall event number
98652

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗