Open original FDA record ↗Recall record
- FDA identifier
- Z-1924-2013
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA classification date
- Aug 7, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Siemens initiated a recall due to confirmed complaints of low patient sample recovery with Dimension TACR Flex Reagent cartridge (DF107) lot BB4087 and its linked calibrator (DC 107) lot 3BD029. Falsely low Tacrolimus results across the assay range may lead to alterations in dosing.
Source facts
- Action
- Siemens sent an Urgent Medical Device Recall letter dated July 2013 to affected customers The letter identified the affected product, problem, potential risk, and actions to be taken by customers. Customers were requested to discuss the letter with thier laboratory director to determine if additional follow up is appropriate for patient tests conducted with the affected device lots. Additionally, customers were asked to discard any remaining inventory and to complete and return the field correction effectiveness check for replacement product. They were also asked to forward the letter to anyone wha had received this affected product. For questions contact your Siemens Technical Solutions Center or your local Siemens Technical support representative.
- Initiated
- 2013-07-01
- Root cause
- Process control
- Terminated
- 2014-08-26
- Distribution
- USA Nationwide Distribution including the states of: TX, NC, GA, TN, UT, NJ, MI, VA, AR, LA, MO, FL, AZ, CA, DC, PA, NH, WI, SD, SC, NM, MD, and IN.
- Recall number
- Z-1924-2013
- Recall status
- Terminated
- CDRH record ID
- 119941
- FDA classified
- 2013-08-07
- Recall event number
- 65706