Skip to content

Public FDA record

Z-1924-2024: ClearCut S Safety Sideport Knife 1.0mm Dual Bevel Model/Catalog Number: 8065771540 Software Version: N/A

Recall record Z-1924-2024 for Alcon Research LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1924-2024
Company in FDA record
Alcon Research LLC
FDA classification date
May 30, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Corneal/Scleral V-Lance Ophthalmic knives.

Source facts

Action
On April 22, 2024 VOLUNTARY MEDICAL DEVICE FIELD CORRECTION letters to customers. Actions to be taken by the Customer / User: If a knife does not perform as expected, or if you feel it does not meet your expectation for sharpness, discontinue use of that specific knife and replace the knife with a new one to avoid potential complications. To acknowledge your receipt of this Voluntary Medical Device Correction notification, please take the following steps: 1. Forward this notification to all departments or organizations using Alcon Ophthalmic Knives. 2. Follow the risk mitigation precautions provided in this notice when using identified catalogue numbers of ophthalmic knives. 3. Please complete the attached Response Form indicating your understanding of the included instructions and return the attached Response Form via email or fax to Alcon. Email: Market.Actions@alcon.com Fax: 817-302-4337 Alcon has reported this issue to Health Authorities in accordance with applicable regulations. In the event you have experienced adverse events or product quality issues related to this communication, please contact Alcon via web (https://notifeye.alcon.com), by email (msus.safety@alcon.com) or by phone (1-800-757-9780). Adverse events or quality problems experienced with the use of this product may also be reported to the FDA MedWatch Adverse Event Reporting program either online (FDA EMAIL), by regular mail or by fax: http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm. Should you have any questions or concerns about this matter or need help finding a replacement or substitute sterile standalone ophthalmic surgical knife, please call Alcon Customer Service or contact your Alcon Sales Representative.
Initiated
2024-04-22
Root cause
Process control
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Columbia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, India, Ireland, Italy, Japan, Korea, Kuwait, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, Puerto Rico, Romania, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom. Product distributed as sterile stand-alone knives have been placed on HOLD. Product included within Alcon Custom Pak (surgical procedure pack) are continuing to be distributed as ceasing distribution would cause immediate market shortage and result in canceled surgeries.
Recall number
Z-1924-2024
Recall status
Open, Classified
CDRH record ID
207328
FDA classified
2024-05-30
Recall event number
94501

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗