Z-1924-2026: Flamingo Funnel Small, Model Number SQ20012-01

Recall record Z-1924-2026 for SurgiSmoke Solutions. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1924-2026
Company in FDA record
SurgiSmoke Solutions
FDA classification date
Apr 22, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Certain lots of product were associated with a manufacturing nonconformance involving use of a non-gamma sterilization compatible drape.

Source facts

Action
An URGENT: MEDICAL DEVICE RECALL notification letter was sent to customers beginning 3/16/26. Actions to be taken by the Customer/End User: 1. Immediately discontinue use of this product. Note: The lots listed in this notice are the only lots impacted. All subsequently built product contains a correct, gamma compatible drape. 2. Dispose of impacted product according to local and national waste disposal procedures. Note: ALL PRODUCT MUST BE ACCOUNTED FOR ON THE MEDICAL DEVICE RECALL RETURN RESPONSE ACKNOWLEDGEMENT AND RECEIPT FORM. 3. Acknowledge receipt of this notice. Complete and return the attached Medical Device Recall Return Response Acknowledgement and Receipt Form via phone, US Mail, or email even if you no longer have any affected product in inventory: Phone: Communicate the information required in the attached form by calling the toll free number: 1-800-557-3698 - available M-F between 8:00AM and 6:00PM CST US Mail: Complete the attached form, and mail it back using the enclosed self-addressed, stamped envelope (if notice received by mail) Email: Complete the attached form and return it to quality@surgismokesolutions.com Please forward this notification to any additional customers to whom the affected product may have been distributed. Type of Action by the Company: In addition to this Recall activity, Nascent Surgical has initiated Corrective Action and Preventive Action measures that include but are not limited to: Correcting the Bill of Materials to include a gamma compatible drape (completed Jan 2025) Changing the packaging configuration to remove the fold (completed Jan 2025) Improving Quality System Planning and other procedures to prevent recurrence Other Information: If you have additional questions about this notice, please contact us: By phone 1-800-557-3698 (toll free) - available M-F between 8:00AM and 6:00PM CST By emailing quality@surgismokesolutions.com
Initiated
2026-03-16
Root cause
Packaging change control
Distribution
Worldwide - US Nationwide distribution in the states of CA, CO, GA, MA, MD, ME, MN, NH, NY, OH, TN, TX, and UT. The country of Canada.
Recall number
Z-1924-2026
Recall status
Open, Classified
CDRH record ID
219098
FDA classified
2026-04-22
Recall event number
98574

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗