Open original FDA record ↗Recall record
- FDA identifier
- Z-1925-2012
- Company in FDA record
- Maquet Cardiovascular Us Sales, Llc
- FDA classification date
- Jul 6, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class I
Reason stated in the source
A technical alarm may be generated on the FLOW-i system when using the MAN/AUTO switch to change ventilation modes from manual to automatic or from automatic to manual if the switch is not fully engaged in the "on" or "off" position but remains in the "in-between position.
Source facts
- Action
- The firm, Maquet Getinge Group, sent an "URGENT DEVICE FIELD CORRECTION" letter dated February 20, 2012 to its customers. The letter describes the product, problem and action to be taken. The letter state that a Maquet Service Representative will upgrade the customers software in the FLOW-i Anesthesia System to the current revision and also provide the customers with the corresponding user's manual. Note: This upgrade includes other minor improvements. The customers were instructed to sign a document to verify satisfactory completion of the work once the software upgrade was completed. If you have questions or require additional information, please contact your local MAQUET representative or Technical Support at 888-627-8383. A press release was issued on 7/6/2012.
- Initiated
- 2012-02-20
- Root cause
- Device Design
- Terminated
- 2015-09-09
- Distribution
- Worldwide distribution: USA (nationwide) state of GA only and countries of: Argentina, Australia, Austria, Belgium, China, Czech Republic, Ecuador, Finland, France, Germany, India, Ireland, Israel, Italy, Latvia, Netherlands, Norway, Poland, Portugal, Romania, Russia, Serbia, Singapore, South Africa, Spain, Sweden incl Maquet Critical Care, Switzerland, Thailand, United Arab Emirates, and United Kingdom.
- Recall number
- Z-1925-2012
- Recall status
- Terminated
- CDRH record ID
- 109875
- FDA classified
- 2012-07-06
- Recall event number
- 62118