Surface finish is different than specified. The implant has 30 grit blast on the exterior of the flange when the print calls for the glass bead blast. The patient may experience pain requiring surgical intervention.
Source facts
Action
Biomet sent an Urgent Medical Device Recall Notice letter dated May 28, 2014 to their customer. The letter identified the affected product, problem and actions to be taken. The letter instructed physician to monitor the patient for any pain that may require surgical intervention. For questions call (574) 372-1570.
Initiated
2014-05-15
Root cause
Nonconforming Material/Component
Terminated
2015-03-16
Distribution
US Nationwide Distribution in the state of California
Recall number
Z-1925-2014
Recall status
Terminated
CDRH record ID
127549
FDA classified
2014-06-26
Recall event number
68418
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.